VEGA WEIGHTRAC 31 4 … 20 mA_HART - four-wire User Manual

Page 53

Advertising
background image

53

7 Setup with PACTware

WEIGHTRAC 31 • 4 … 20 mA/HART - four-wire

42374-EN-131119

You can select further options in PACTware.

A detailled description of the functions is available in the PACTware

online help.

When a reset is carried out, all settings (with only a few exceptions)

are reset. The exceptions are: PIN, language and HART mode.
The following reset functions are available:
Basic settings: Restores the parameter settings at the time of ship-

ment Ex factory including the order-specific settings. The measured

value memory is deleted.
Factory settings: Resets parameter settings including special

parameters to the default values of the respective instrument. The

measured value memory and the order-specific settings are deleted.
Peak values: Resets the parameter settings in the menu item "Setup"

to the default values of the respective instrument. Order-specific

settings remain but are not taken over into the current parameters.

The measured value memory as well as the event memory remain

unaffected.
The following table shows the default values of the instrument. The

values apply for the application "Mass flow". The application must be

selected first.
Depending on the instrument version, not all menu items may be

available or they may be differently assigned:

Menu

Menu item

Default value

Setup

Measurement

loop name

Sensor

Isotope

Cs-137

Application

Mass flow

Background ra-

diation

0 ct/s

Unit of the pro-

cess value

m

Temperature unit °C
Damping

60 s

Real value cor-

rection

0

Current output

mode

4 … 20 mA, < 3.6 mA

Current output

Min./Max.

Min. current 3.8 mA, max. current

20.5 mA

X-ray alarm

Modulated measuring current

Reference value

- Relay

None

Block adjustment Released

Further settings in PACT-

ware

Additional adjustments/

Reset

Advertising