VEGA VEGABAR 87 Modbus and Levelmaster protocol - Operating Instructions User Manual

Page 9

Advertising
background image

9

3 Product description

VEGABAR 87 • Modbus and Levelmaster protocol

46297-EN-131107

The VEGABAR 87 is a pressure transmitter for pressure and level

measurements of liquids with higher temperatures in the chemical,

food processing and pharmaceutical industry.

Measured products are liquids.
Depending on the instrument version and the measurement setup,

the measured products can be also viscous.

The process pressure acts on the sensor element via the stainless

steel diaphragm and an internal transmission liquid. The process

pressure causes a resistance change which is converted into a cor-

responding output signal and outputted as measured value.
The METEC

®

measuring cell is the measuring unit. It consists of

the ceramic-capacitive CERTEC

®

measuring cell and a special,

temperature-compensated isolating system.

4

3

2

1

Fig. 3: Configuration of the METEC

®

measuring cell with VEGABAR 87

1 Process diaphragm

2 Isolating liquid

3 FeNi adapter

4 CERTEC

®

measuring cell

The measuring cell design depends on the selected pressure type.

Relative pressure: the measuring cell is open to atmosphere. The

ambient pressure is detected in the measuring cell and compensated.

It thus has no influence on the measured value.
Absolute pressure: the measuring cell is evacuated and encapsu-

lated. The ambient pressure is not compensated and does hence

influence the measured value.
Relative pressure, climate-compensated: the measuring cell is

evacuated and encapsulated. The ambient pressure is detected

through a reference sensor in the electronics and compensated. It

thus has no influence on the measured value.

The following illustration shows the installation of the CERTEC

®

measuring cell in the transmitter as well as the sealing concept.

Application area

Measured products

Measuring system

Pressure types

Seal concept