VEGA VEGAFLEX 82 Modbus and Levelmaster protocol User Manual

Page 3

Advertising
background image

3

Quick start

VEGAFLEX 82 • Modbus and Levelmaster protocol

41833-EN-130314

1. In this menu item you can select the application.

2. In the menu item "Medium - Dielectric constant" you can define

the medium type (medium).

3. Carry out the adjustment in the menu items "Min. adjustment" and

"Max. adjustment".

4. A "Linearization" is required for all vessels in which the vessel vol-

ume does not increase linearly with the level - e.g. in a horizontal

cylindrical or spherical tank. Activate the appropriate curve.

5. A "False signal suppression" detects, marks and saves the false

signals so that they are no longer taken into account for level

measurement. We generally recommend a false signal suppres-

sion.

The sensor measures the distance from the sensor (reference plane)

to the product surface.

4

5

3

1

2

4

1

2

3

5

Fig. 3: Measuring ranges - VEGAFLEX 82
1 Reference plane

2 Probe length L

3 Measuring range

4 Upper dead band (in this area no measurement is possible)

5 Lower dead band (in this area no measurement is possible)

For this adjustment, the distance is entered when the vessel is full

and nearly empty. If these values are not known, an adjustment with

other distances, for example, 10 % and 90 % is also possible. Starting

point for these distance specifications is always the seal surface of

the thread or flange.

Parameterization example

Advertising