VEGA VEGAFLEX 82 Modbus and Levelmaster protocol User Manual

Page 33

Advertising
background image

33

6 Set up with the display and adjustment module

VEGAFLEX 82 • Modbus and Levelmaster protocol

41833-EN-130314

In the menu item "Current output Min./Max.", you determine the reac-

tion of the current output during operation.

The default setting is min. current 3.8 mA and max. current 20.5 mA.

Since the HART variables are very extensive, the indication was

divided into two menu items.
In the menu "HART variables 1" you can have the first HART value PV

(primary value) and the second HART value SV (secondary value) of

the sensor displayed.
The values cannot be changed in the display and adjustment module.

To change these values, you have to use the adjustment software

PACTware.

In the menu "HART variables 2" you can have the third HART value

TV (third value) and the fourth HART value QV (quarter value) of the

sensor displayed.
The values cannot be changed in the display and adjustment module.

To change these values, you have to use the adjustment software

PACTware.

The following circumstances cause interfering reflections and can

influence the measurement:

High sockets

Vessel installations such as struts

Setup - Current output

Min./Max.

Setup - HART variables

(1)

Setup - HART variables

(2)

Setup - False signal sup-

pression

Advertising