ProLab 4000

Ion concentration

119

ba75646e03

12/2012

5

Thoroughly rinse the electrode with deionized water.

6

Immerse the electrode in the sample.

7

Start the measurement with [Continue].
After end of the measurement, the Standard pops up.

8

In the Standard register, enter the value for Standard concen-
tration
if necessary.

9

Use [Dosing] to automatically add to the sample the adjusted
volume of the standard solution (Standard volume) (1 vol% of
sample solution).
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (Standard volume) (1 vol% of sample solution).