ORP voltage

89

ba75646e03

12/2012

8

ORP voltage

8.1

General information

You can measure the following variables:

 ORP [mV]

 Relative ORP [mV]

Temperature

measurement

The temperature value is measured and documented with all measure-
ments.
You have the following options to measure the temperature:

 Automatic temperature measurement with

– Temperature sensor (NTC30 or Pt1000) integrated in electrode

(if available).

– the integrated temperature sensor of another electrode that is

connected and immersed in the sample at the same time.

– an external temperature sensor (NTC30 or Pt1000).

 Manual determination and input of the temperature.

The display of the temperature indicates the active temperature mea-
suring mode:

Temperature

sensor

Resolution of
the temp. dis-
play

Mode

yes

0.1 °C

Automatic with temperature
sensor

-

1 °C

Manual