Xylem ProLab 4000 User Manual

Page 155

Advertising
background image

ProLab 4000

Conductivity

155

ba75646d03

12/2012

Note
The measuring method can only be started if Ref. temp., Start temp.
and Temp. interval are selected so that the reference temperature cor-
responds to the nominal temperature of a measuring point: Ref. temp.
= Start temp. + x*Temp. interval

5

Enter the start temperature, end temperature and temperature
interval.
Enter the number of standards and confirm with [Continue].
The display shows:
 the current conductivity value
 the current temperature
 information on the nominal temperature of the next measur-

ing point and the temperature difference between the cur-
rent temperature and the nominal temperature

6

Warm or cool the measuring solution (standard 1) to the start
temperature Start temp..
As soon as the temperature of the measuring point is achieved
and a stable value is recognized, the measurement is automat-
ically started and the dataset stored.
After the measurement, a piece of information about the next
measuring point with the relevant nominal temperature is dis-
played and the temperature difference between the current
temperature and the nominal temperature.