3 conductivity measurement – Xylem ProLab 4000 User Manual

Page 200

Advertising
background image

What to do if...

ProLab 4000

200

ba75646e03

12/2012

Warning [TempErr]

Warning [ISEErr]

Burette does not dose

16.3

Conductivity measurement

Conductivity measuring

cell was not recognized

Error message, Error

(invalid calibration)

Cause

Remedy

– Temperature difference

between measurement and cali-
bration greater than 2 K.

– Adjust the temperature of

the test sample

Cause

Remedy

– Electrode voltage outside cali-

brated range

– Recalibrate

Cause

Remedy

– Address of the burette incor-

rectly set

– Enter the address of the

burette on the meter
(see page 101)

– Enter the address of the

burette on the burette
(see operating manual of the
burette)

– RS232 interface of the burette

incorrectly set

– Configure the RS232 inter-

face of the burette
(see page 29 and operating
manual of the burette)

Cause

Remedy

– Measuring cell not connected

– Connect measuring cell

– Cable broken

– Replace measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable control standard

– Use control standard,

0.01 mol/l KCl