ProLab 4000

Commissioning

29

ba75646e03

12/2012

3.7.1

RS232 interface (serial port)

PC, printer and burettes can be connected via the RS232 interface.

Data that is transmitted to the interface with <PRINT> can be printed
out with a printer or received with a terminal program if a PC is con-
nected.

Operation with RS232

Connect the interface of the ProLab 4000 to the devices using the Z390
(PC) or Z893 (external USB printer Z890) cable.

Set up the following transmission data on the PC/printer:

Note
If the burette control is active, data output to the RS232 interface
(printer) is deactivated.

On the meter, the following transmission data are permanently set for
the RS232 interface:

– Baud rate: 4800

– Data bits: 8

– Stop bits: 1

– Parity: none

The transmission data must agree with the transmission data for the
interface adjusted on the burette.
Check the transmission data adjusted on the burette, and change
them if necessary (see operating manual of your burette).

Baud rate

selectable between: 1200 ... 115200
The baud rate must agree with the baud rate set on
the PC/printer.

Handshake

RTS/CTS

PC only:

Parity

none

Data bits

8

Stop bits

1

Note
Pin assignment of the socket: See page 203.