Model mt 3810 – Brooks Instrument MT3810 User Manual

Page 12

Advertising
background image

Model MT 3810

Installation and Operation Manual

X-VA-MT3810-eng

Part Number: 541B050AAG

December, 2008

Section 1 Introduction

1-6

Specifications - SMM Microprocessor Transmitter with or without Alarm
and Pulse Output

EMC Directive 89/336/EEC: EN 50081, EN 50082 and EN 61326-1

Hazardous Location Classification
Enclosure: Type 4X/ IP65
Ambient Temperature: -22°F> Tamb < 150°F (-30°C > Tamb < 65°C)

Intrinsically Safe
United States and Canada UL Listed, E73889, Vol. 1, Sect. 15
Class I, II and III, Division 1, Groups A, B, C, D, E, F, and G; T4

Europe – KEMA 01ATEX1235 X

II 2 G EEx ia IIC T4
II 2 D T135

o

C

Entity Parameters (Transmitter):
Ui=Vmax=30 Vdc; Ii=Imax=140 mA; Ci= 15 nF; Li= 0 mH

Entity Parameters (Integral Alarms):
Ui=Vmax=30 Vdc; Ii=Imax=45 mA; Ci= 0 nF; Li= 0 mH

Non-Incendive
United States and Canada UL Listed, E73889, Vol. 1, Sect. 15
Class I, II, III, Division 2, Groups A, B, C,D F, and G; T4

Europe – KEMA 01ATEX1236

II 3 G EEx nA II T4
II 3 D T135

o

C