3-10
Model MT 3810
Installation and Operation Manual
X-VA-MT3810-eng
Part Number: 541B050AAG
December, 2008
Section 3 Operation
From the Measurements Menu,
select
Flow.
From the Flow Menu, select
FlowRate Units of Measure
(FlowRate UOM).
In Flow, scroll through list
optional units. Highlight the
FlowRate UOM desired. Press
ENTER
(F4) to enter the new
information, and press
SEND
(F2) to send this information to
the transmitter.
Note: Two Warning messages will appear.
•
Warning Pressing OK will change device output. Put loop in manual.
•
Warning Return control loop to automatic control.
These warnings confirm changes to the programming and alert the user to
make changes in other related fields.
F1
F4
F3
F2
SAVE
Brooks MT 38xxVA
Measurements
5 Viscosity
1 Flow
2 Temperature
3 Density
4 Pressure
F1
F4
F3
F2
SAVE
Brooks MT 38xxVA
Flow
1 FlowRate Units
2 Configuration
3 Meter Calibration
F1
F4
F3
F2
ENTER
Brooks MT 38xxVA
FlowRate Units
gal/min
L/min
I mpgal/min