2-3

Installation and Operation Manual
X-VA-MT3810-eng
Part Number: 541B050AHG
June, 2008

Model MT 3810

Section 2 Installation

2-7 Installation of Flowmeter

If the inlet and outlet valves adjacent to the indicator are to be
closed for any reason, the indicator must be completely drained.
Failure to do so may result in thermal expansion of the liquid
which can rupture the meter and cause possible personal injury.

WARNING

Recommended installation for Models MT 3809 and MT 3819 is as
follows:

a. Carefully remove the covers from each end of the flowmeter.
b. Install the flowmeter with the inlet at the bottom and the outlet at the

top.

c. When installing the flowmeter in the process line, follow accepted

plumbing practices for flanged or NPT fittings.

d. Install the flowmeter within 5° of true vertical. Use of a level is

recommended to determine the proper alignment.

e. Installation of a bypass piping arrangement is recommended,

Figure 2-1. Bypass piping permits the meter to be isolated from the
flow line for servicing and cleaning.

Figure 2-1 Typical Bypass Installation

Advertising