Model mt 3810 – Brooks Instrument MT3810 User Manual

Page 40

Advertising
background image

3-2

Model MT 3810

Installation and Operation Manual

X-VA-MT3810-eng

Part Number: 541B050AAG

December, 2008

Section 3 Operation

Transmitter Only (uP)

Transmitter (uP) Plus...

Pre-Programmed Parameters

Alarms and Pulse Output
Preprogrammed Parameters

Model Number

Alarm LoLimit (0% FS flow)

Serial Number

Alarm HiLimit (100% FS flow)

Tag Name (blank)

Pulse output units (1 ppg)

Flow Rate units of measure (gpm)

Resettable totalizer units of measure (gal.)

Low Flow cutoff (10% FS flow)

Inventory totalizer units of measurement

Calibration factor (% scales only)

(gal.)

Analog Output LoRange

(4 mA @ 0% flow)

Analog Output HiRange

(20 mA @ 100% flow)

c. Manual Programming or Reprogramming of the SMM electronics

If electronic parameters were not preprogrammed prior to shipment or
new settings are required, programming of the SMM transmitter with or
without alarms and pulse output may also be done locally via an
Emerson Model 275 hand held Communicator. Refer to Figures 3-1
and 3-2 for Device Description Basic Setup and Detailed Device
Description Setup. Trees generate an electrical signal directly
proportional to the gas mass flow rate.