Fs10 button controls – Fluid Components International FS10A User Manual

Page 17

Advertising
background image

FS10 Series

Fluid Components International LLC

15

FS10 Button Controls

It is recommended the unit be powered-up for 10-15 minutes before making changes to any of the flow settings.
General Control Description
The FS10 control “functions” can be accessed through the two buttons by:

1. Entering “function selection”
2. Selecting the desired function to adjust
3. Adjusting the function control
4. Exiting to normal operation

1. To Enter “Function Selection” mode:

a. Press and hold both buttons down for 3 seconds
b. When “function selection” is entered, all the LEDs will flash “on”, then
c. The first function (LED # 1) will be illuminated

2. “Function Selection” mode:

Once the unit is in “function selection” mode, the buttons take on new controls:

• The “-“ and “+” buttons will step (increment down or up respectively) through the different functions. The current function number will

be indicated on the LED blinking at the slow rate.

• Holding either the “+” or “-“ for 3 seconds will “select” the current function to be adjusted.

3. When a Function has been selected for adjustment:

After selecting the desired function by pushing and holding either the “-” or “+” button for 3 seconds the LEDs will flash at a faster rate. The

pattern of the LEDs will either reflect the current value of the parameter being adjusted, or the “ready to capture” pattern for parameters to be

captured (see specific function descriptions for details).

The buttons will either increment or decrement the function parameter, or capture a value for the corresponding parameter.

4. To exit out of the current function, push and hold both buttons down for 3 seconds. The LEDs will indicate that you are once again in “function

election” mode, and the current mode will have incremented to the next function. To exit completely (into operational mode), press the two

buttons simultaneously for 3 seconds once again.

Note:

Function #4 (load a bank) and function #15 (save to a bank), requires holding either button for 3 seconds to load or save.

Holding both buttons will exit either function without performing the operation.

Button Operation

1

Enter “function selection”

Push and hold both buttons simultaneously 3 seconds

2

Select desired “function”

Push “+” or “-“ to step through (LEDs indicate function #)

3a.

Enter “function adjust”

Push and hold either “-“ or “+” button 3 seconds

3b.

Adjust within function

Follow instructions for particular function entry

4a.

Exit “function adjust”

Push and hold either “-” or “+” button 3 seconds (now in “function selection”

operation) [NOTE: Function #4 or #15 requires holding both buttons to exit into function

select operation]

4b.

Exit “function selection”

Push and hold both buttons simultaneously 3 seconds (now in normal operation)

Table 3

Advertising