Normal set-up and operation using the button interface

Refer to Table 5, Button Controls to address the following functions in the field.
Bank Selection of Stored Parameters (Function 4)
Select function 4 to change bank selection of stored parameters. The FS10 has 7 storage banks. Banks 1-6 are factory set defaults of common

applications or specific factory calibrations. Bank 7 is available for user saved parameters. Refer to complete FS10 configuration number to

determine the factory supplied bank setting. Note, an error pattern will result when selecting through the banks if the bank contains no valid

calibration parameters.
Units that do not come factory calibrated use Universal Setting A from Bank 1 as the settings for operational Bank 0. These parameters cover

the entire range (delta R) of the device. Therefore, spans and trip-points can be set in the field using this bank regardless of process fluid.

Once a unit is set-up in the field, the changed parameters now in Bank 0 may be stored to bank 7 using Function 15 and later retrieved using
Function 4 if unwanted changes were made in the interim

Bank Default Values

Bank No.

Process Fluid

Range Setting

0

FS10 Current Active Parameters

1

Universal Default Setting A - Low flow

sensitivity

Un-calibrated output - low sensor excitation power setting

Full range gas or liquid

2

Air/Gas -40 °F to 250 °F (1/4 ″ tube and SP76

adapter)

1 SFPS to 50 SFPS (200 cc/min to 10,000 cc/min, [0.4 SCFH to 20 SCFH] in

1/4 “ tube) – trend accuracy 10% full scale

3

Universal Setting B - High flow gas,

liquids

Un-calibrated output – 1.5X sensor excitation power setting

Full range liquid and gas

4

Water/Glycol -40 °F to 250 °F

0.02 SFPS to 0.5 SFPS (4 cc/min to 100 cc/min [0.001 GPM to 0.03 GPM] in

1/4” tube) – trend accuracy 10% full scale

5

Reserved for custom calibration

6

Reserved for custom calibration

7

User defined and stored

Table 4