FS10 Series

4

Fluid Components International LLC

2

INSTRUMENT INSTALLATION

The FS10 is marked with a flow direction arrow or “A” etched onto the sensor element. It is located on the flattened area of the sensor body

close to the housing or on the assembled tee. In 1/4 and 3/8 inch tube tees, the “A” should be mounted upstream of the flow to maximize sensitivity

at low flow rates (flow into the “A”). Larger line sizes should follow the flow arrow direction. Refer to Table 1 for optimum results. Where the flow

tube is not included with the sensor assembly, the orientation mark must be parallel to flow (±3°). For liquid vertical flows in particular, FCI

recommends the sensor element be installed where flow is in the up direction. In vertical low flow gas applications, flow in the

down direction is recommended.
As a level device, the orientation mark can be perpendicular or parallel to the liquid level. The sensor element may be installed top mount 90° to

the liquid surface. The sensor element can be at any angle as long as the flow direction follows the flow arrow. Liquid applications where the flow

element is positioned other than horizontally, FCI recommends the flow go in the up direction.

CAUTION: To minimize the possibility of damage, leave the protective covers over the sensing area until the time of installation. Take

extra precaution with the sensing elements and surface when installing.

For NPT process connections, apply the appropriate sealant compatible with the process media to the male threads. Tighten until the orientation

mark is positioned correctly. Check for leaks.

Special Conditions for Safe Use:

a. Provisions shall be made to prevent the rated voltage being exceeded by transient disturbances of more than 40%.
b. For applications in explosive atmospheres caused by air/dust mixtures, cable and conduit entries used shall provide a degree of ingress

protection of at least IP 54 according to EN 60529.

c. When the temperature under rated conditions exceeds 158 °F [70 °C] at the cable or conduit entry point, or 176 °F [80 °C] at the branching

point of the conductors, the temperature specification of the selected cable shall be in compliance with the actual

measured temperature.

d. Cable gland assemblies are factory tightened – do not adjust; M12 connectors assembled finger-tight. Mencom MDC-8MR-PG9 or equal

M12 connector is used on FS10A.

1

INSTRUMENT DESCRIPTION AND IDENTIFICATION

The FS10A is a universal flow monitor and switch specifically designed for gas and liquid process analyzer sampling systems. The FS10A is a fast

responding, highly repeatable sensor which installs easily into a standard tube tee fitting or SP76 (NeSSI) modular manifold.