Command Line Interface Commands

INFO

Read version information (Hardware, Firmware, Date code, other values (#1.. #45). [The values in items

1 through 45 are listed]

MEAS

Read most recent flow measurement values (#220..#238). [The values in items 220 through 238 are listed]

RCFG B

Read configuration parameters (#80..#132) from bank B (0..9). [Example: *RCFG 7 (to read bank 7 parameters)]

SAVE B

Save currently active parameters to bank B (0..9) (Level 2 – Field password required). [Example: *SAVE 7]

RCL B

Recall parameters from bank B as the active parameters. (Level 2 – Field password required). [Example: *RCL 7]

PASSWD NNNN

To enter the password for either Level 1 (Factory) or Level 2 (Field). [Example *PASSWD 19113]

EXIT

To exit the current password level (undo having entered a password). The current parameters will automatically be

saved to Bank 0 upon exit.

NNN=<value>

Update item #NNN (1..238) in the “active” (RAM) parameters with value. Value may be either integer or floating

point (scientific notation). [Example 1: *228=1.100119E+03 enters value 1.100119E+03 into item 228, Example 2:

*1=002 enters 002 into item1, Example 3: *110=800 enters 800 into item 110]

B:NNN

Read item #NNN (1..238) in Bank B (0..9). For items #1..#45, and items #220..#238 the Bank is ignored. If no bank is

specified, the currently active (RAM) parameters will be accessed. [Example 1: *7:85 returns item 85 from bank 7 in

this form: 7:85>5.053665E-02, Example 2: *8:119 returns item 119 from bank 8 in this form: 8:119=030]