Feed rate parts replacement – Hydrotech Classic Series Stenner Basic Service Manual User Manual

Page 34

Advertising
background image

33

Feed Rate Parts Replacement

1. Remove screws.

2. Remove feed rate mounting plate and dial ring.

3. Remove cam, spider assembly, and index plate.

If required, remove index pin assembly by unscrewing with
box wrench and replace.

4. Flip worn index plate over OR install new index plate.

5. Clean out FRC channel before installing new cam.

6. Use grease (AquaShield

®

) to lubricate the cam’s angled tip.

7. Install the cam.

8. Feed the angled tip into the FRC channel and keep the

90° end in the up position. Place thumb over the
cam to guide as it is inserted in the channel.

9. Continue to feed the cam in the channel until there is

approximately a 1/2" gap between the cam tip and the
beginning of the channel

.

Installing
New Cam

Continued...