Appendix A Technical Specification A-17

Table A-10 ADAM-4018/4018+ Range Accuracy for

Thermocouple

Input

Range

Code

(Hex)

Input Range

Typical

Accuracy

Maximum

Error

Units

0E

J thermocouple
0 to 760 ℃

±1.0 ±1.5

0F

K thermocouple
0 to 1370 ℃

±1.0 ±1.5

10

T thermocouple
-100 to 400 ℃

±1.0 ±1.5

11

E thermocouple
0 to 1000 ℃

±1.0 ±1.5

12

R thermocouple
500 to 1750 ℃

±1.2 ±2.5

13

S thermocouple
500 to 1750 ℃

±1.2 ±2.5

14

B thermocouple
500 to 1800 ℃

±2.0 ±3.0


NOTE:
Because the CJC sensor of ADAM-4018 is located in
the side of channel 0 to 4, the measurement will have the
difference ± 1 ℃ between channel 0 ~ 4 and channel 5 ~ 7.