Luminex xPONENT for MAGPIX IVD User Manual

Page 11

Advertising
background image

1. Open the Maintenance page, then the Cmds & Routines tab.

2. Select Revive After Storage (Luminex) from the drop-down list. The Revive After

Storage routine performs the following commands:

Prime

Rinse

Alcohol Flush

Rinse

3. On the Cmds & Routines tab, select Prime from the Commands option section and

perform a Prime.

4. Add 70% isopropanol or ethanol to reservoir RB1 on the off-plate reagent block as

indicated on the Cmds & Routines tab.

NOTE: The rinse reservoir (RD1) should be empty.

5. Click Retract.

6. Click Run.

FIGURE 2.

Revive After Storage (Luminex) Routine

After the Revive After Storage routine is complete, run the System Initialization routine.

For In Vitro Diagnostic Use

Starting the MAGPIX System for the First Time