When you turn on the Luminex MAGPIX system, the following start up indicators occur:

• Blue light indicator turns on.

• Fans turn on.

• Syringe pump initializes.

• The filter wheel initializes.

• Software connects to the instrument.

NOTE: It may take several minutes for the instrument to initialize and

connect to the software.

The Software Connectivity indicator is located on the lower left
corner of the software screen.

Log On to xPONENT

If your version of xPONENT is licensed for 21 CFR Part 11 and/or Security, an application
administrator must set up user IDs (and passwords, if required). If you are not using a version
with 21 CFR Part 11 and/or security, users can log in with any user name or with no user
name.

CAUTION:

Use of this software by untrained personnel can result in
inaccurate data and test results. Personnel who will use
xPONENT must read this manual thoroughly before operating
the software.

1. On the System Login tab, type your user ID.

2. If you are using a secure version of the software, type your password. The Home page

opens.

Initial Startup

When you turn on the system for the first time, perform the following procedures.

1. Adjust the Sample Probe Height

2. Revive After Storage (Luminex) Routine

3. System Initialization

• Fluidics Preparation

• Calibration

• Performance Verification

Adjusting the Sample Probe Height

Adjust the sample probe height to ensure that the probe drops far enough into the well to
acquire sample.

NOTE: Ensure that there is no liquid in the wells or reservoirs before

adjusting the sample probe height.

MAGPIX

®

User Quick Guide

2

For In Vitro Diagnostic Use