Problem Summary

Resolution

Notes

If Calibration continues to fail run
the following routine:

Create routines in the
xPONENT “Maintenance” Tab
“Cmds & Routines” page
Approx run time: Five mins

(1x) Clean with 0.1 N NaOH
(reservoir C)

(3x) Rinse (reservoir D)

Run Calibration Routine

Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.

If problem is not solved, contact Luminex Technical Support

Performance Verification
Failing

Sonicate and flush the sample
probe. Observe the probe
calibration for positional
accuracy. Verify all fittings are
secured in place (reference the
MAGPIX Calibration and
Performance Verification Trouble
Shooting Guide
flowchart below).

Refer to the MAGPIX
Calibration and Performance
Verification Trouble Shooting
Guide
flowchart below for
detailed information on the
fittings.

If system continues to fail run the
enhanced start-up routine
outlined below:

Create routines in the
xPONENT “Maintenance” Tab,
“Cmds & Routines” page
Approx run time: Thirty mins

Sonicate and flush the sample
probe (flush from narrow to wide
end).

(4x) Rinse (reservoir D)

(3x) Alcohol Flush with 70%
isopropanol or ethanol alcohol
(reservoir B)

(1x) Alcohol Flush with DI water
(reservoir A)

For In Vitro Diagnostic Use

Troubleshooting