Creating a Kit

To create a kit:

1. Open the Protocols page, then open the Protocols tab. Select the protocol that you

want to use for the kit, then click NewStd/Ctrl. The Std/Ctrl Details tab opens.

2. Click on the protocol that you want to use for the kit, then click New Std/Ctrl. The Std/

Ctrl Details tab opens.

3. Type the name of the kit in the Name box, the lot number in the Std/Ctrl Kit Lot# box,

the expiration date using MM/DD/YY format in the Expiration box, and the manufacturer
in the Manufacturer box.

4. Click Apply Std Lot if you want to apply a standard lot. The Select Lot dialog box opens.

Click a lot and select OK.

5. Click Apply Ctrl Lot to apply a control lot. The Select Lot dialog box opens. Select a lot

and click OK.

6. Alternatively, type the appropriate information in the Assay Standard Information and

Assay Control Information sections. The number of standards and/or controls in these
sections is defined in the protocol. If your batch uses controls, select Expected, Low or
High from the Show Value options. Use the Apply Values arrows to apply values down
or across the range of analytes.

7. Click Save.

Creating a Lot

To create lots, you must use a protocol that uses either Quantitative or Qualitative analysis
settings.

To create a lot: