22

Operating up to four Headstages rotatory with one Wireless-System

It is possible to operate up to four different headstages connected to one Wireless-System
in a rotatory system. The headstages are discriminated via frequency band for the radio
communication. Each headstage is labeled with a letter (A, B, C, D). Please see the bill of delivery
for the assignment of the headstage and the frequency band. (When using one single headstage
only, the headstage is automatically A without a label.)

After pressing the button “Scan” the receiver scans which headstages are available.
Then the user is able to select the desired headstage.

After recording from the first headstage, you can choose the second headstage in the “Wireless”
window, the third and the fourth.

Operating up to four Wireless-Systems in parallel with one Computer

It is also possible to operate up to four different complete Wireless-Systems with one data
acquisition computer. This way, you can experiment with up to four laboratory animals
contemporaneous. The headstages are discriminated via frequency band for the radio
communication again. Headstages and related receivers are labeled for differentiation.

Set up the Wireless-Systems and connect them to the data acquisition computer.
Start as many instances of MC_Rack software as you need. In “Data Source Setup”
assign the Wireless-Systems via serial number to the instances of MC_Rack.