Multichannel Systems MEA1060-Up-BC Manual User Manual

Page 35

Advertising
background image

Troubleshooting

35

If the List Mode is open, there is an additional button Trace List.

If the Trace List button is active all other functions are automatically disabled. The stimulation
patterns of the list, created in List Mode, are applied one after the other as stimulation pulse.
To control which pattern of the list is reached the name of the stimulus pattern will be
highlighted in blue. Reaching the end of the list, the program reruns the list of the stimulation
patterns, starting at the top of the list again. It is not recommended to change the settings of
stimulation patterns during experiment. Please stop recording data, change settings and send
the new information to the amplifier with Download. Then start recording again. It is possible
to save the list of stimulation patterns by clicking Save under Configuration together with a
previously loaded configuration file or you can save the list only. The file for just the list has the
same extension as the configuration file ".scl".

Note: A Firmware update is needed to use this feature.

During the blanking signal, the stimulus inputs are connected to the software-selected
stimulating electrodes, and all MEA electrodes are disconnected from the amplifier.
The last signal value on all electrode channels before the blanking pulse is applied will be
saved and then kept until the channels are reconnected to the amplifier.