Time of day

The hour will fl ash. Press  to increase
and  to decrease hour. Press  to
advance to minutes. Once minutes are
fl ashing, press  to increase or  to
decrease minutes. Press  to move to
next parameter.

CLr

Press

[] to reset accumulated

run time and volume from pump
memory to zero. See below.

ESC

Press

[] to exit User Interface

Level 2 without saving changes to
parameters.

End

Press

[] to save changes to

parameters and exit User Interface.

OFF

Appears only when a DataTrac 2000
program or delayed start is loaded in
pump memory. When OFF displays,
press [] to delete and exit the
User Interface. See page 15.

Resetting Run Time Data

To reset accumulated volume and run time data to zero, enter User Interface
Level Two by placing a running pump in HOLD and pressing 
within 10 seconds. Press  until CLr appears on the LCD. Press [] to
clear the run time data. Press  until End appears, and then press [].
The pump is now in HOLD.

CLr does not clear previously set sample time (SГ). See Deleting a
Sample Time on page 16.

Setting Pump Flow Rate

1. With the pump in HOLD, press [] to run the pump and enter

 within 10 seconds. You are now in User Interface Level One.

2. The

fl ow rate on the LCD will fl ash. Press  to increase fl ow or  to

decrease fl ow until the desired fl ow rate is displayed.

3. Press

 repeatedly until End appears on the LCD.

4. Press [] to save the fl ow rate and exit the User Interface.
5. Press [] again to place the pump in HOLD if desired.