Calibration – SKC 210-2000 AirChek 2000 User Manual

Page 21

Advertising
background image

18

9. Press  until End appears on the LCD. Press [] to save fl ow sett ing

and exit the User Interface. The pump will remain in RUN. Reset run
time data (see page 13).

10. Place pump in HOLD by pressing []. Disconnect the calibration

train.

Pressing [] will place a running pump in HOLD, and a holding
pump in RUN.

If a delayed start or DataTrac 2000 schedule has been programmed
into the pump, it may remain in pump memory. PROG will display
in the upper left corner of the pump display. To delete this program,
enter User Interface Level Two, scroll to the fl ashing OFF, and press
[
].

Calibration