4

Table 1. AirChek 2000 Run Time in Hours with

NiCad Battery

Following are typical run times achieved when using a fully charged nickel-

cadmium (NiCad) battery pack. Data is sorted by type of sample media. All run

times are listed in hours. Results obtained using a new pump and new fully
charged battery. Pump performance may vary.

Note

Mixed Cellulose (MCE) fi lter, 0.8-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

24.2

15.2

2.5

20.4

12.4

3.0

17.7

**

Polyvinyl Chloride (PVC) fi lter, 5.0-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

28.2

21.8

2.5

27.0

22.0

3.0

22.6

18.2

** Filter back pressure exceeded pump capability during testing.

Increases in back pressure during sampling due to buildup of
sample on the fi lter can decrease batt ery life.

Table 2. AirChek 2000 Run Time in Hours with

NiMH Battery

Following are typical run times achieved when using a fully charged nickel-

metal hydride (NiMH) battery pack. Data is sorted by type of sample media. All

run times are listed in hours. Results obtained using a new pump and new fully
charged battery. Pump performance may vary.

Note

Mixed Cellulose (MCE) fi lter, 0.8-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

37

32

2.5

33

26

3.0

30

21

Polyvinyl Chloride (PVC) fi lter, 5.0-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

44

40

2.5

36

34

3.0

31

27

Increases in back pressure during sampling due to buildup of
sample on the fi lter can decrease batt ery life.