Performance profi le – SKC 210-2000 AirChek 2000 User Manual

Page 6

Advertising
background image

3

Performance Profi le

Time Display:

Time of day in hours and minutes (12 or 24-hour
clock) with AM and PM indicators

Volume Display:

Continually updated, based on corrected fl ow rate
multiplied by sampling time

Flow Fault:

If the pump is unable to compensate for longer than
15 seconds due to excessive back pressure, the pump
enters fl ow fault mode. During fl ow fault, the pump goes
into HOLD, the fault icon appears on the display during
the length of the fault, and the accumulated run time
display is frozen and retained. After 5 minutes in fl ow fault,
auto-restart is attempted every 5 minutes up to 10 times.

Power Supply:

Rechargeable 4.8-V nickel metal hydride (NiMH)
battery pack,
3.5-Ah capacity or

Rechargeable 4.8-V nickel cadmium (NiCad) battery
pack,
2.0-Ah capacity

A

battery eliminator is available (see Accessories);

use voids the UL Listing for intrinsic safety.

Charge Time:

6 to 8.5 hrs using PowerFlex charger

(varies with battery capacity
and level of discharge)

Size:

5.6 x 3 x 2.3 in (14.2 x 7.6 x 5.8 cm)

Weight:

22 oz (624 gm)

RFI/EMI Shielding:

RFI/EMI-shielded case, CE marked

Intrinsic Safety:

UL and cUL Listed

ATEX and ANZEx Models available

Tubing:

Requires 1/4-inch ID tubing

CE marked

UL Listed for intrinsic safety

Use only SKC-approved parts to ensure reliable performance. Failure
to do so voids any warranty and the UL Listing for intrinsic safety.