14

Setting Sample Time (S Г)

Program the AirChek 2000 from the keypad to run STEL, TWA, or any other
run time from 1 to 999 minutes*. AirChek 2000 sampling time may also
be programmed from a PC using DataTrac 2000 Soft ware*. To program a
sample time using the pump keypad:

1. With the pump running, press [] to place

the pump in HOLD and enter  within 10
seconds. You are in User Interface Level Two.

2. Repeatedly

press

 until

and a fl ashing 00

appear on the display.

3. Set the sample time by pressing  to increase it

or  to decrease it to the desired time in minutes.

4. Press

 repeatedly until End appears.

5. Press

[] to save the sample time and exit the

User Interface. A fl ashing S will appear with the
selected sampling time on the LCD.

6. Press

[] to begin sampling. The time display

will count down, the pump will go into HOLD,
and the total sample time will display when
sampling is complete.

7. To delete a set sample time, enter User Interface

Level Two, scroll to

, and press  until time

appears as 00. Exit the User Interface by scrolling
to End and pressing [].

If a time still displays aft er canceling a sample time, this display is
the total run time since the data was last reset.

If a sample time () has been programmed into the pump, a
DataTrac 2000 program cannot be entered without deleting the
sample time fi rst (see page 16
). Likewise, if a DataTrac 2000 program
resides in pump memory, the sample time (
) function cannot be
selected until the DataTrac program is deleted (see page 16
).

* Sample Time () Sett ing: AirChek 2000 sampling time can be set
up to 999 minutes from the pump keypad or up to 43,200 minutes
(30 days) from a PC using DataTrac 2000 Soft ware. During sampling, the
pump LCD will display remaining run time (will count time down from
a set number of minutes) when sample time is set from the keypad.
The pump LCD will show elapsed time in minutes when the pump
is programmed from a PC. If run time is longer than 9999 minutes
(6.8 days), the sample time displayed on the LCD will roll over to
1 aft er reaching 9999.