Setup, Setting a delayed start, Deleting a delayed start – SKC 210-2000 AirChek 2000 User Manual

Page 18

Advertising
background image

15

Setting a Delayed Start

The Delayed Start feature is available in pump version 2.59 or higher.

When sett ing the pump manually for sampling to begin within a 12-hour
timeframe, follow this procedure:

1. With the pump running, press [] to place the pump in HOLD and

enter  within 10 seconds. You are in User Interface Level Two.

2. Press

 until

appears. Press  or  to set a sample time. This is

required to set a delayed start.

3. Press

 until the display reaches the 12-hr/24-hr clock. Press  or 

until the display shows a fl ashing dELA (delayed start).

4. Press

. Time will display with hours fl ashing. Press  to increase or

 to decrease the hour. Press  to move to minutes. Press  to increase
or  to decrease the minutes.

The time entered here will be the next occurrence of this time. There
is no a.m. or p.m. designation.

5. Press

 until End appears.

6. Press

[].

7. The PROG icon and a fl ashing HOLD will appear in the upper left

corner of the display. The pump is now set for delayed start.

If sett ing a delayed start using DataTrac 2000 Soft ware, refer to the
DataTrac 2000 Operating Instructions (included on the soft ware CD).
A programmed delayed start can be viewed in the DataTrac 2000
Scheduler.

Deleting a Delayed Start

This is the same method used to delete a DataTrac 2000 program from the
pump (see Deleting a DataTrac 2000 Program on page 16).

1. With the pump running (will only run in short bursts), press [] to

place the pump in HOLD and enter  within 10 seconds. You are
in User Interface Level Two.

2. Press

 to scroll to the fl ashing OFF indicator and press [].

3. Press

 until End displays.

4. Press

[]. The PROG icon should be gone. If not, repeat process until

it disappears.

Setup