Operating the pocket pump in constant flow mode – SKC Limited Pocket Pump MTX User Manual

Page 18

Advertising
background image

Page 16

210-PMTXM Issue F

www.skcltd.com

Operating the Pocket Pump in Constant Flow Mode

2) Setting and Calibrating the Pump Flow Rate

Ensure that the pump is set to Constant Flow mode and that the battery is fully charged.
Set up the sample train as specified in the sampling method and connect to the pump air inlet port. Connect
a flow meter such as a Defender primary flow calibrator to the inlet of the sample train.

Press [ ] to start the pump running.

Press . The LCD screen will show the flow rate setting screen with ‘

SET

flashing.

Use the and buttons to set the required flow rate in millilitres per minute on the
LCD screen.

When the required flow rate is set press the button to display the flow rate adjust
screen, with ‘

ADJ’ flashing.

Set the flow calibrator to take continuous readings, and observe the indicated flow
rate. Use the and buttons to adjust the pump flow rate up or down, to precisely
calibrate the required flow rate.