14

922251-00 Rev

A

Cartridges

Replaceable Socket Cartridge

If the input tip socket gets damaged, you don’t have to
replace the entire probe, because the HFP series active
probe has a removable tip socket cartridge.

PACC-MS002, package of 1.

High Frequency Cartridge

By having a fixed tip rather than a socket, the High Frequency cartridge is able to increase signal
fidelity at higher frequencies.

The High Frequency Cartridge is an optional accessory for the HFP1000.

PACC-MS003, package of 1.

Removal and Installation of the Replaceable Cartridge

To remove old cartridge:

1. Slide the cable strain relief over the cable away from

the probe body.

2. To release the latch, lift the part closest to the strain

relief away from the probe body and slide the
cartridge toward the strain relief.

To install a new cartridge:

1. Slide the new cartridge onto the probe body until the

latch engages.

2. Slide the cable strain relief forward to cover the back

end of the probe body.

NOTE: The cable strain relief is polarized and fits over the probe body in one direction only.