Preliminary procedure – Teledyne LeCroy HFP1000 User Manual

Page 29

Advertising
background image

Operator’s Manual

922251-00 Rev A

23

Table 2 List of Required Equipment

Description

Minimum Requirements

Example Test Equipment

Digital Oscilloscope

ProBus interface

Teledyne LeCroy WavePro960
Teledyne LeCroy LT344

Digital Multimeter (DMM)
with test probe leads

4.5 digit
DC: 0.1% Accuracy
AC: 0.1% Accuracy

Agilent Technologies 34401A
Fluke 8842A-09

Function Generator

Sine Wave output amplitude
adjustable to 14.14 Vp-p (5
Vrms) into 1 MΩ at 70 Hz

Agilent Technologies 33120A
Stanford Research Model DS340

BNC Coaxial Cable (2 ea.)

Male to Male, 50 Ω, 36" Cable

Pomona 2249-C-36
Pomona 5697-36

BNC Tee Connector

Male to Dual Female

Pomona 3285

Calibration Fixture

ProBus Extender Cable

Teledyne LeCroy PROBUS-CF01

Terminator, Precision, BNC

50 Ω ± 0.05%

Teledyne LeCroy TERM-CF01

Banana Plug Adapter (2 ea.)

Female BNC to Dual Banana
Plug

Pomona 1269

BNC to Mini-grabber

BNC Male to Mini-grabber
Cable, 36"

Pomona 5187-C-36

Preliminary Procedure

1. Connect the HFP1000 probe to the female end of the ProBus Extension Cable. Connect

themale end of the ProBus Extension Cable to channel 1 of the oscilloscope.

2. Turn the oscilloscope on and allow at least 30 minutes warm-up time for theHFP1000 and

testequipment before performing the Verification Procedure.

3. Turn on the other test equipment and allow these to warm up for the timerecommended by

the manufacturer.

4. While the instruments are reaching operating temperature, make a photocopy ofthe

PerformanceVerification Test Record (located in Appendix A), and fill in the necessary data.

5. Select the channel to which the probe is connected. Set the oscilloscope scalefactor to

20mV/div.