Teledyne LeCroy HFP2500 User Manual

Page 33

Advertising
background image

Operator’s Manual

922253-00 Rev A

27

7. Record the DMM reading, which should be a negative number, to 10 mV resolution as

’Power Supply Negative Output Voltage’ in the Test Record.

8. Add 10 (to correct for the +10 V offset as described in step B-13) to the ’PS Negative Output

Voltage’ recorded in step B-7. (Do NOT adjust the power supply output amplitude).

9. Divide the resulting sum by 10.

10. Record the answer to three significant places as ’Expected Negative Output Voltage" in the

test record.

11. Remove the banana plug adapter, connected to the power supply, from the DMM and

connect the precision 50 Ω terminator to the DMM, making sure that the banana plug side
marked ’GROUND" is connected to the LOW or COMMON input of the DMM.

12. Set the DMM to read DC Volt on the most sensitive range.

13. Verify that the display for channel 1 is turned ON. Turn the oscilloscope OFFSET knob to

read +10.00 V on the oscilloscope display.

14. After the DMM has settled, record the reading to 100 μV resolution as ’Measured Negative

Output Voltage’ in the Test Record.

15. Subtract the measured voltage as recorded in step B-14 from the expected output voltage

recorded in step B-10. Be sure to include the sign of each of the values in the calculation.

16. Record the answer to three significant places as ’Offset Error Voltage’ in the Test Record.

17. Verify that the error is between ±10.8 mV.

NOTE: The error term is derived from the Offset Accuracy specification of ±1% ±8 mV. Using
a 10.0 V offset setting, the maximum error would be 108 mV referred to the input, which
becomes ±10.8 mV error referred to the output (taking into account the ÷10 attenuation).

18. Using the oscilloscope’s OFFSET knob, set the probe offset to 0 V, as indicated in the

onscreen display.

19. Remove the dual banana plug adapter with the HFP2500 attached from the power supply

and reconnect to the supply but now with the grounded side of the banana plug (and
grounded socket of the probe head) connected to the negative terminal of the power
supply output.

20. Disconnect the DMM from the precision 50 Ω terminator and connect the DMM to the dual

banana plug adapter connected to the power supply output.

21. Record the DMM reading, which should be a positive number, to 10 mV resolution as

’Power Supply Positive Output Voltage’ in the Test Record.

Advertising