Operator’s Manual

922253-00 Rev A

29

8. After the DMM reading has stabilized, record the reading to 100 μV resolution as ’Measured

Output Voltage, top range’ in the Test Record.

9. Calculate the error by dividing the measured output voltage recorded in step C-8 by the

expected top output voltage recorded in step C-6. Subtract 1 from this ratio and multiply by
100% to get the error in percent.

10. Record the calculated error to two decimal places (±0.xx%) as ’Gain Error, top range’ in the

test record.

11. Verify that the error is less than ±1.0 %.

12. Disconnect the precision 50 Ω terminator from the DMM.

13. Connect the banana plug adapter connected via a BNC cable to the BNC tee at the function

generator to the DMM. Verify that the side of the plug marked ’Ground’ is connected to the
LOW or COMMON input of the DMM.

14. Adjust the sine wave generator output amplitude to approximately 2.5 Vrms as measured

on the DMM.

15. Record the reading to 1 mV resolution as ’Generator Output Voltage, mid range’ in the Test

Record. Be careful not to alter the output amplitude after the reading is recorded.

16. Divide the reading recorded in step C-15 by 10.

17. Record the result to 100 μV resolution as ’Expected Output Voltage, mid range’ in the test

record.

18. Remove the banana plug adapter from the DMM and connect the precision 50 Ω terminator

to the DMM, making sure that the banana plug side marked ’GROUND" is connected to the
LOW or COMMON input of the DMM.

19. After the DMM has stabilized, record the reading to 100 μV resolution as ’Measured Output

Voltage, mid range’ in the Test record.

20. Calculate the error by dividing the measured output voltage recorded in step C-19 by the

expected top output voltage recorded in step C-17. Subtract 1 from this ratio and multiply
by 100% to get the error in percent.