Hfp2500 high frequency probe, Hfp2500 test record – Teledyne LeCroy HFP2500 User Manual

Page 46

Advertising
background image

HFP2500 High Frequency Probe

40

922253-00 Rev

A

HFP2500 TEST RECORD

Step

Description

Intermediate Data

Test Result

Output Zero Voltage

A-5

Output Zero Voltage (Test limit

≤± 800 μV)

__________ V

Offset Accuracy

B-7

Power Supply Negative Output Voltage

__________ V

B-10

Expected Negative Output Voltage

__________ V

B-14

Measured Negative Output Voltage

__________ V

B-16

Offset Error Voltage (Test limit

≤± 10.8 mV)

_________mV

B-21

Power Supply Positive Output Voltage

__________ V

B-23

Expected Positive Output Voltage

__________ V

B-26

Measured Positive Output Voltage

__________ V

B-28

Offset Error Voltage (Test limit

≤± 10.8 mV)

_________mV

LF Attenuation Accuracy

C-5

Generator Output Voltage

__________ V

C-6

Expected Output Voltage, top range

__________ V

C-8

Measured Output Voltage, top range

__________ V

C-10

Gain Error, top range (Test limit

≤± 1.0%)

__________%

C-15

Generator Output Voltage

__________ V

C-17

Expected Output Voltage, mid range

__________ V

C-19

Measured Output Voltage, mid range

__________ V

C-21

Gain Error, mid range (Test limit

≤± 1.0%)

__________%

Advertising