Teledyne LeCroy HFP2500 User Manual

Page 34

Advertising
background image

HFP2500 High Frequency Probe

28

922253-00 Rev

A

22. Subtract 10 from the output voltage recorded in step B-21. Divide this number by 10.

23. Record the result to three significant places as ’Expected Positive Output Voltage’ in the Test

Record.

24. Set the oscilloscope OFFSET to –10.00 V as read on the oscilloscope display.

25. Remove the banana plug adapter from the DMM and connect the precision 50 Ω terminator

to the DMM, making sure that the banana plug side marked ’GROUND" is connected to the
LOW or COMMON input of the DMM.

26. Record the DMM reading to three significant places as ’Measured Positive Output Voltage’

in the Test Record.

27. Subtract the Measured Output Voltage as recorded in step B-26 from the Expected Output

Voltage recorded in step B-23. Be sure to include the sign of the values in the calculation.

28. Record the result to 100 μV resolution as ’Offset Error Voltage’ in the Test Record.

29. Verify that the output error is between ±10.8 mV.

30. Return the oscilloscope offset to 0 Volt. Leave the setup connections for the next step.

C. LF Attenuation Accuracy

1. Disconnect the BNC tee at the power supply from the dual banana plug adapter. Connect

the BNC tee to the output of the function generator. (Use a 50 Ω termination if the function
generator requires such a load).

2. Disconnect the DMM from the precision 50 Ω terminator and connect the DMM to the dual

banana plug adapter connected to the function generator output.

3. Set the DMM to read AC Volt and set the range to measure 5.0 Vrms.

4. Set the mode of the function generator to sine wave, the frequency to 70 Hz and the output

amplitude to 5 Vrms ±10 mV as measured on the DMM.

5. Record the output voltage to 1 mV resolution as ’Generator Output Voltage’ in the Test

Record. Be careful not to alter the output amplitude after the reading is recorded.

6. Divide the reading recorded in step C-5 by 10 and record the result with 100 μV resolution

as ’Expected Output Voltage, top range" in the Test Record.

7. Remove the banana plug adapter, connected to the function generator, from the DMM and

connect the precision 50 Ω terminator to the DMM, making sure that the banana plug side
marked ’GROUND" is connected to the LOW or COMMON input of the DMM.