Manual dial – ZOLL E Series Monitor Defibrillator Rev B User Manual

Page 11

Advertising
background image

12-Lead Reports

9650-1218-01 Rev. B

9

Use the scroll keys on top of the unit to select a number
or character on the keypad, then press the Commit ( )
key on top of the unit to enter the digit in the Phone #
field. Repeat for each additional digit (up to 25
characters).

When you have entered the phone number, select the
Enter (

 ) key from the keypad and press the

Commit ( ) key on top of the unit to initiate dialing, or
press the Dial softkey.

Note: If at least one digit has been entered, pressing

the Dial softkey or highlighting the Enter (

) key

on the keypad and pressing the Commit ( ) key
dials the entered phone number and displays the
following screen:

Standard modem call status messages appear as the
various phases of the transmission progress.

When transmitting to a ZDR server, the TRANSMITTING
message is displayed while transmission is in progress.
A successful transmission will display the message
TRANSMISSION COMPLETE.

24

NIBP

-

---

---

SpO2%

---

CO2

---

m

m

Hg

---

RR

ECG

96

II x1.5



kPa



Dial

Cancel

2

3

4

6

,

7

8

9

+

#

*

0

Phone #

|

5

w

x

›››

‹‹‹

1

w

MANUAL DIAL

24

NIBP

-

---

---

SpO2%

---

CO2

---

m

m

Hg

---

RR

ECG

96

II x1.5



kPa



Dial

Cancel

2

3

4

6

,

7

8

9

+

#

*

0

Phone #

5

w

x

›››

‹‹‹

1

w

MANUAL DIAL

16175551234

Abor t

Alar ms

When transmitting to a FAX machine, the following
messages display in sequence: FAX PREPARING, FAX
DIALING, FAX SENDING, FAX DONE
.

For direct FAX only: If you need to abort the
transmission, press the Abort softkey. Note that if you
turn the selector switch to another mode of operation
(DEFIB or PACER), the transmission is automatically
aborted.

The Retry screen is displayed whenever one of these
things occur:

The transmission is complete.

An error has occurred.

The Abort softkey is pressed.

Pressing Retry displays the Transmission screen, so
that transmission can be performed again. Pressing
Return displays the main 12-lead screen. If 30 seconds
elapse and a softkey is not pressed, the main 12-lead
screen will be displayed.

For ZDR server transmissions, the E Series unit
automatically retries the transmissions up to 3 times. In
this case, the message RE-TRANSMITTING is
displayed.

Note the following fault conditions:

Transmission Error

In the event of a failure to transmit the ECG record,
you are notified with a screen message and tones,
and the E Series unit continues to retry the
transmission until the operator aborts the
transmission or the transmission is successful. See
Troubleshooting on page 20 for possible
transmission error messages and the corrective
actions you can take.

Check Recorder

If the recorder runs out of paper or a paper jam
occurs during printout of the 12-lead report, you are
notified with a screen message and tones, and data
transmission continues. The operator can retrieve the
12-lead report by printing a summary report after the
error condition is cleared.

Retr y

Alar ms

Retur n