12

9650-1218-01 Rev. B

Patient Information

Measurements Matrix

You can configure the unit to print out a measurements matrix, which consists of measurements for each segment of
each lead. To print the measurements matrix at the end of an analysis, you must change the default setting in the
E Series device. (For more details, see the E Series Configuration Guide.)

12SL Wave Measurement

After the P, QRS, and T complexes have been demarcated, the waves for each complex are identified. This is done
separately for each lead. The program finds the points at which the signal crosses the baseline within each complex. If
the crossing points define a wave that has an area greater than or equal to 160

Vms, the wave is considered to be

significant. If the area is less than 160

Vms, the program considers the wave to be insignificant and does not label it

as a separate wave. The measurement matrix contains the amplitudes (with respect to QRS onset) and durations of all
of these individual waves.

DEVICE ID: 00000000000

RECORDED: 16:09:27 31 MAY 05

PATIENT NAME : JOE SMITH

PATIENT ID: 200505311609

PATIENT AGE: 60

PATIENT SEX: Male

Vent. Rate : 60

PR interval 166 ms

QRS Duration: 82 ms

QT/QTc 358 / 358 ms

P-R-T axes 54 43 47

Normal sinus rhythm
Normal ECG

*** Unconfirmed ***