Symptom recommended action – ZOLL E Series Monitor Defibrillator Rev B User Manual

Page 24

Advertising
background image

E Series - 12 Lead ECG Monitoring Option Insert

22

9650-1218-01 Rev. B

While attempting to pair a Bluetooth device
with the E Series unit, the NO DEVICES
FOUND
message is displayed.

Verify that your Bluetooth device is turned on.

Verify that your Bluetooth device is within 30 feet of the E Series
unit.

Verify that the correct device from the list of discovered devices
has been selected.

Verify that your Bluetooth device is in discoverable mode.

Repeating the pairing will often succeed.

Note:

Some devices will automatically turn off discoverable mode
after a very short period; check your device’s documentation.

When adding additional Bluetooth devices to
the list of recipients, the LIST MEMORY FULL
message is displayed.

The E Series unit can only pair up to 5 Bluetooth devices.
Remove one or more devices before attempting to pair any
additional ones.

When adding additional ZOLL Data Relay
servers to the list of servers, the LIST
MEMORY FULL
message is displayed.

The E Series unit can only send data to up to 12 ZDR servers.
Remove one or more servers before attempting to add any
additional ones.

When adding additional contacts to the Contact
List, the LIST MEMORY FULL message is
displayed.

The E Series unit can only support up to 36 contacts. Remove
one or more contacts before attempting to add any additional
ones.

When attempting to delete a server from the
list, the SERVER RECORD USED BY
CONTACT LIST, UNABLE TO DELETE
message is displayed.

A server cannot be deleted if there are contacts using that server.
Remove all contacts before attempting to delete the server.

When attempting to PING a Bluetooth device
from the E Series unit, the PING TEST FAILED
message is displayed.

Verify that your Bluetooth device is turned on.

Verify that your Bluetooth device is within 30 feet of the E Series
unit.

Verify that your device has its Bluetooth functionality enabled.

Verify that the server and Bluetooth device configurations are
correct.

Note:

Cell signals vary depending on the carrier and area.

When attempting to transmit a 12-lead report,
the TRANSMISSION FAULT XX message is
displayed.

There may be an internal problem with the E Series unit; contact
ZOLL technical support.

Symptom

Recommended Action

Advertising