E Series - 12 Lead ECG Monitoring Option Insert

20

9650-1218-01 Rev. B

Troubleshooting

The troubleshooting section is intended to help you identify and correct problems that arise during operation. If trouble
persists after consulting this information, contact the ZOLL Technical Service Department for problems relating to the
E Series unit.

ZOLL Technical Service Department

(USA) Phone: (800) 348-9011

(UK) Phone: +44-192-584-6400

(Other locations) Contact your local ZOLL distributor.

Symptom

Recommended Action

General

The ECG baseline does not stabilize, the ECG
is noisy, or the NOISY ECG message is
displayed.

Verify the electrodes are connected properly to the patient.
Reposition the electrodes and/or lead wires to prevent the
electrodes from pulling away from the patient.

Verify the gel of the electrode is not dry. Check expiration date on
the electrode package.

Use Silver/Silver Chloride electrodes.

Verify that the patient cable is connected properly and held
motionless.

Verify that the ECG snaps and connectors are clean.

Make sure that the patient is motionless. Support patient’s limbs,
if necessary.

Stop vehicle while acquiring the 12-lead ECG.

Verify correct filter (50 or 60 Hz) is selected in Configuration
menu.

Inspect ECG cable. Replace the cable if damaged.

Change ECG filter setting to 0.05-40 Hz 4x3, and retry
acquisition.

ECG LEAD OFF or ECG VX LEAD OFF
message is displayed.

Confirm proper ECG electrode and cable connections.

Prepare skin and replace electrodes.

Check the ECG cable continuity and replace the cable if you
suspect it is damaged.

Note: ECG signal may be temporarily out of range due to recent

defibrillator discharge.

DATA NOT READY / SEND FROM PAT. REC.
message is displayed.

12-lead data is not yet ready for transmission. Select the 12-lead
record from the Patient Records screen and retransmit.