UIDE

18

Set Lead Groups

This setting allows the user to configure two custom lead groups as described in the 12-Lead

ECG Monitoring insert (part number 9650-0215-01). Pressing the Change Value softkey will

cause the following screen to appear:

Pressing the Prev Lead softkey will select the previous lead.
Pressing the Next Lead softkey will select the next lead.
Pressing the Change Lead softkey will change the selected lead. Repeatedly pressing this key

will cycle the selected lead through all twelve possibilities.
Pressing the Enter softkey will return to main configuration menu.

12 Lead Freq Response

When set to “0.05-150 Hz 4x3” the 12-Lead Filter Setting defaults to 0.05-150 Hz 4x3.
When set to “0.05-150 Hz Cont.” the 12-Lead Filter Setting defaults to 0.05-150 Hz Cont.
When set to “0.05-40 Hz 4x3” the 12-Lead Filter Setting defaults to 0.05-40 Hz 4x3.
Refer to the 12- Lead ECG Monitoring Operator’s Guide insert (part number 9650-0215-01)

for details on each setting.

Lead Group Default

When set to “Standard” the 12-Lead Group Setting defaults to Standard.
When set to “Custom Group 1” the 12-Lead Group Setting defaults to the setting defined under

Custom Group 1.
When set to “Custom Group 2” the 12-Lead Group Setting defaults to the setting defined under

Custom Group 2.
Refer to the 12- Lead ECG Monitoring Operator’s Guide insert (part number 9650-0215-01)

for details on each setting.

Print 12 Lead Interpretation

When set to “Yes” the 12SL interpretative statements will print following the acquisition of a

12-lead ECG. This setting affects the strip chart and summary report printouts as well as the fax

transmission.

Defib Default to Pads

When set to “Yes” switching the unit to DEFIB mode will set the active lead to Pads/Paddles

regardless of the configuration setting for “Display Lead on Power-Up.”