Cell phone type, Etco2 options, Enable leads off – ZOLL M Series CCT Defibrillator Rev A User Manual

Page 23: Blood pressure units, Nibp auto interval default, Trigger nibp meas on bp alarm, Trigger nibp meas on hr alarm, Nibp first inflate pressure default, Hone, Co2 o

Advertising
background image

Configuration Settings

19

Cell Phone Type

This setting controls the active cellular phone profile for PCMCIA modem cards. It has no

effect on landline functionality. Pressing the Change Value softkey selects different phone

makes.
Select the phone make corresponding to your target cellular phone. For additional information

on cellular phone compatibility contact The Supply Net, Inc. at

www.thesupplynet.com

or

(800) 826-0279.

EtCO

2

Options

Pressing Set/Review causes a submenu to display, which allows review of or changes to the

following capnography options:

Default EtCO2 Units:

Sets EtCO

2

units to mmHg, kPa, or %

Displayed Zoom Level:

Sets the default scale setting for the displayed capnographic
waveform. Does not affect the printed waveform or the
waveform data stored on the data card.

Default EtCO2 Comp Setting: Sets the default compensation for CO

2

Enable Leads Off

When set to “Yes” and ECG leads do not have proper contact with the patient, or the cable is not

properly attached to the M Series CCT unit, the ECG LEADS OFF message displays and a

dashed line displays in place of the ECG waveform. Yes is the default setting.
Note:

When using 12 Lead or Pacer mode, the M Series CCT unit functions as if this

configuration were set to “Yes.”

When set to “No” and ECG leads do not have proper contact with the patient or the cable is not

properly attached to the M Series CCT unit, the POOR LEAD CONTACT message and the

ECG waveform display.
Caution: Use care when interpreting ECG waveforms when the POOR LEAD CONTACT

message displays, because the ECG waveform may include a significant amount of

artifact. Whenever possible, reattach the leads before interpreting the waveform.

Blood Pressure Units

Sets the NIBP and IBP unit of measurement to mmHg or kPa.

NIBP Auto Interval Default

Sets the default for the amount of time between automatic measurements.

Trigger NIBP Meas on BP Alarm

When set to “Yes,” the M Series CCT unit initiates a single additional blood pressure

measurement when any of the NIBP alarms trigger.

Trigger NIBP Meas on HR Alarm

When set to “Yes,” the M Series CCT unit initiates a single blood pressure measurement when

heart rate/pulse rate alarm triggers.

NIBP First Inflate Pressure Default

Sets the default NIBP cuff inflation pressure (the pressure up to which the NIBP cuff inflates at

the beginning of each measurement).