Basic auto energy escalation, Asic, Nergy – ZOLL M Series CCT Defibrillator Rev A User Manual

Page 24: Scalation, 6\vwhp&rqiljxudwlrq

Advertising
background image

M S

ERIES

CCT C

ONFIGURATION

G

UIDE

20

On M Series CCT units, you can set the NIBP cuff inflation pressure for three patient type

modes: adult, pediatric and neonatal. With NIBP First Inflate Pressure Default highlighted and

Set/Review selected, press the Change Value softkey. The following screen displays:

Pressing the Change Value softkey will scroll through the available inflation pressures for the

highlighted parameter.
Pressing the Prev Item or Next Item softkeys moves the highlight to the previous or next

parameter in the menu.
Pressing the Return softkey selects the values that have been entered and returns to the main

configuration menu.

Basic Auto Energy Escalation

When set to Yes, the unit automatically increments the defibrillation energy to the levels

specified in Energy Level: Shock 1, 2, 3 after each of the first two shocks and displays the

message “ENERGY INCREMENTED,” when both of the following are true:

The defibrillator is in Manual mode

The defibrillator has external paddles or pads attached

Manually changing the energy level outside the pre-programmed sequence and delivering a

shock disables this function until the unit is turned off for more than 10 seconds, then turned

back on.
This option increments the energy level regardless of Shock Advisory activation or status. It

does not work with internal handles.
For more information about the energy level settings, see the “Energy Level: Shock 1, 2, 3”

section of this manual.