Advisory and semi-automatic modes, Dvisory, Automatic – ZOLL M Series CCT Defibrillator Rev A User Manual

Page 35: Odes

Advertising
background image

Configuration Option Tables

31

Advisory and Semi-automatic Modes

Feature

Options

Default

Energy Level: Shock 1

Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J

Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J

Manual DSW: 200 J
Manual Biphasic: 120 J

AED DSW: 200 J
AED Biphasic: 120 J

Energy Level: Shock 2

Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J

Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J

Manual DSW: 300 J
Manual Biphasic: 150 J

AED DSW: 200 J
AED Biphasic: 120 J

Energy Level: Shock 3

Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J

Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J

Manual DSW: 360 J
Manual Biphasic: 200 J

AED DSW: 360 J
AED Biphasic: 200 J

Auto Analyze 3 Times

Yes, No, Only in Semi-Auto

Manual: No

AED: Only in Semi-Auto

“NO SHOCK ADVISED” Prompt

“NO SHOCK ADVISED”, “CHECK PULSE”,
BOTH

“NO SHOCK ADVISED”

“CHECK PATIENT” Prompt

“CHECK PATIENT”, “PRESS ANALYZE”

“CHECK PATIENT”

Display “DO CPR” Message

Yes, No, Only in Semi-Auto

Manual: No

AED: Only in Semi-Auto

Display ECG in Semi-Auto Mode

Yes/No

Yes

Store to Card in Auto Monitor Mode

Yes/No

Yes

Auto Charge in Advisory Mode

Yes/No

Yes

Record to Card in Advisory Mode

None, ECG Only, Voice and ECG

None

Enable Voice in Advisory Mode

Yes/No

No

Display Heart Rate in Semi-Auto Mode

Yes/No

No

Auto-Analyze on Power Up

Yes/No

No

Enable 12-Lead in Semi-Auto Mode

Yes/No

Yes

AED Defib Ready Hold Time

15 seconds, 60 seconds

15 seconds

Advertising