Trending enabled, Battery type used, Rending – ZOLL M Series CCT Defibrillator Rev A User Manual

Page 26: Nabled, Aed d, Efib, Eady, Aed s, Utomatic, Attery

Advertising
background image

M S

ERIES

CCT C

ONFIGURATION

G

UIDE

22

every 5, 10, 15, 30, or 60 minutes. All data samples are printed on the strip chart regardless of

this setting.

Trending Enabled

When set to “Yes”, the M Series CCT unit will print and display trend reports as long as the

appropriate hardware is installed. Defaults to “No” for units not equipped with the appropriate

hardware.

AED Defib Ready Hold Time (AED Semi-Automatic
Mode)

Sets the number of seconds the unit sounds a charge ready tone, indicating that the defibrillator

is charged and ready to use. The charge ready tone stops when you discharge the defibrillator or

when the hold time for the charge ready tone has elapsed. You can set the hold time for the

charge ready tone to 15 seconds or 60 seconds. The default is 15 seconds.

Battery Type Used

Allows the user to select between the PD™ 4410 battery (default) and the newer XL™ battery.

To maximize battery life, it is important that the setting reflect the actual battery in use. Each

battery type has a different threshold for “low battery” and “shutdown”, and has a different

Amp-hour rating.

WARNINGS:

If an XL battery is installed in an M Series CCT, the user must ensure that the “BATTERY

TYPE USED” is set to “XL Battery.” Otherwise, the M Series CCT may abnormally

terminate battery charging prior to completion when charging a depleted or nearly depleted

XL battery. If battery charging is abnormally terminated, the M Series CCT battery will

toggle the “CHARGER ON” light between amber and green, and, when M Series CCT

power is applied, it will display the message “BATT. OVERCHARGE” on the screen for

approximately 30 seconds.

Note that if battery charging is abnormally terminated because the “BATTERY TYPE

USED” has been incorrectly set to “PD 4410”, the XL battery can still be charged to

completion by cycling M Series CCT power from “off” to “on” and then back to “off” to

clear this condition.

Incorrectly setting “BATTERY TYPE USED” to “PD 4410” when an XL battery is installed

will cause the “LOW BATTERY” warning message to be displayed on the M Series CCT

screen prematurely.

Incorrectly setting the selection to “XL battery” when a PD 4410 battery is installed will

inappropriately delay issuance of the “LOW BATTERY” warning and severely limit the

number of defibrillation shocks which can be delivered between the onset of the “LOW

BATTERY” warning message and the M Series CCT shutdown.