General

Feature

Options

Default

Selected Language

English, Other

English

Notch Filter (ECG)

50, 60 Hz

60 Hz

Allow Card Erase

Yes/No

Yes

Display Elapsed Time

Yes/No

AED: Yes

Manual Advisory: No

* Voice Markers Enabled

Yes/No

No

Set Report Restart Delays

Set/Review

Set/Review

Summary Report Restart Delay

5, 15, 30, 90 minutes, 1.5 days

15 minutes

Trend Report Restart Delay

5, 15, 30, 90 minutes, 1.5 days

15 minutes

Print 12 Lead 4x3

None, 1 Copy, 2 Copies

None

* Auto Self-Test TIme-out Period

No Test, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17,
18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29,
30 days

No test

Async Softkey in Pace Mode

Yes/No

Yes

Code Markers
(Defib Code Markers, Monitor Code
Markers, Pace Code Markers)

Config/Review

Config/Review

Device Identifier

0 through 9, A through Z, blank, and “_” inclusive
for each digit

0 0 0 0 0 0 0 0

0 0 0

Manual Mode Passcode

0 through 3 inclusive, for each digit

0 0 0

Configuration Mode Passcode

0 through 3 inclusive, for each digit

0 0 0 0 0 0 0 0

Display Lead on Power-Up

Paddles/Pads, I, II, or III

Paddles/Pads

Frequency Response

0.5 - 21 Hz, 0.5 - 27 Hz, 1 - 21 Hz

0.5 - 21 Hz

QRS Volume on Power-Up

Midrange, OFF

Midrange

Pace Rate Setting on Power-Up

30 through 102 ppm (increments of 4 ppm)

70 ppm

Displayed Implanted Pace Spikes

Yes/No

Yes

Gain on Power-Up (ECG)

AGC, 0.5 cm/mV, 1 cm/mV, 1.5 cm/mV, 2 cm/mV

AGC

Report Memory Card Errors

Yes/No

Yes

Enable Leads Off

Yes/No

Yes

Upload Baud Rate

9600 bps, 38400 bps, 115200 bps

38400 bps

Trending Report Zoom Level

All, 5 min., 10 min., 15 min., 30 min., 60 min.

All

Trending Enabled

Yes/No

Yes

Battery Type Used

PD 4410, XL Battery

PD 4410

Alarm Limits at Power-Up

Set/Review

Set/Review

Alarms Active at Power-Up

Yes/No

No

Low Battery Tone

1, 5 minutes

1 minute