Defib/Pacer Parameters

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

109

Pediatric Defib Int. Pdl. Energy

The Pediatric Defib Int. Pdl. Energy parameter allows you to set the energy level of the internal
paddle for pediatric patients.

The values for the Pediatric Defib Int. Pdl. Energy parameter are:

Neonate Defib Int. Pdl. Energy

The Neonate Defib Int. Pdl. Energy parameter allows you to set the energy level of the internal
paddle for neonate patients.

The values for the Neonate Defib Int. Pdl. Energy parameter are:

Sync After Cardioversion

The Sync After Cardioversion parameter allows you to enable or disable synchronizing the
defibrillator discharge with the ECG R-wave.

The values for the Sync After Cardioversion parameter are:

Default Pacer Mode

The Default Pacer Mode parameter allows you to set the default Pacer Mode.

The values for the Default Pacer Mode parameter are:

Demand: pacing pulses are inhibited by the patients QRS complexes that occur during an
interval that is dependant on the setting of the rate control. If during that interval no QRS
complexes are detected, a pacing pulse is delivered to the patient, the number of which and the
rate are selected in the pacing rate window.

Fixed: pacing pulses are not dependant on the patient’s cardiac activity and are delivered at the
selected pacing rate.

Default Pacer Rate

The Default Pacer Rate parameter allows you to set the default Pacer rate.

The values for the Default Pacer Rate parameter are:

Possible Values:

1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J, 10 J, 15 J, 20 J, 30 J, 50 J,

Default:

10 J

Possible Values:

1 J, 2 J, 3 J, 4 J, 5 J, 6 J, 7 J, 8 J, 9 J, 10 J, 15 J, 20 J, 30 J, 50 J

Default:

10 J

Possible Values:

On, Off

Default:

Off

Possible Values:

Demand, Fixed

Default:

Demand

Possible Values:

30-180; <100 by 5’s, >100 by 10’s

Default:

70