ZOLL X Series Monitor Defibrillator Rev B User Manual

Page 117

Advertising
background image

ECG Parameters

9650-001358-01 Rev. B

ZOLL X Series Configuration Manual

111

12-Lead Format

The 12-Lead Format parameter allows you to specify the print format of 12-Lead data.

The values for the 12-Lead Format parameter are:

Print 10s Ldll

The Print 10s Ldll parameter allows you to enable or disable the X Series unit printing 10
seconds of the Lead II waveform after printing a 12-Lead snapshot.

The values for the Print 10s Ldll parameter are:

12-Ld Frequency Response

The 12-Ld Frequency Response parameter allows you to specify the 12-Lead waveform
frequency response.

The values for the 12-Ld Frequency Response parameter are:

Diagnostic: frequency ranges are 0.05-150 Hz.

Filtered Diagnostic: frequency ranges are 0.25-40 Hz.

12-Ld Analysis Enabled

The 12-Ld Analysis Enabled parameter allows you to enable or disable the 12-Lead
Interpretive Analysis capability.

Note:

This parameter is available only if the 12-Lead Interpretive Analysis option has been
purchased.

The values for the 12-Ld Frequency Response parameter are:

Possible Values:

2x6, 3x4, 3x4 Cabrera

Default:

3x4

Possible Values:

On, Off

Default:

On

Possible Values:

Diagnostic, Filtered Diagnostic

Default:

Filtered Diagnostic

Possible Values:

On, Off

Default:

On